SDI12 Box QC

Not Connected

1. Connect Uploader Dongle

Plug the SDI12 uploader dongle into USB, then click Connect.

2. Check the Box

The box takes the standard probe's calibration, stores it and reads the probe at rest. GOOD: pot it. BAD: do what the screen says and click again.

3. Check Firmware Before Shipping

SHIP: it carries the production firmware. UPDATE FIRST: click UPDATE THIS BOX (about 13 minutes over the bus for a box on v2.2 or older, about a minute from v2.3 on), then check again. DO NOT SHIP: potted, the box is out of stock for good — the verdict line says what it may still be used for; not yet potted, back on the Flasher and check again. A potted box cannot be updated, and the 1.3 image resets itself every 19 seconds and goes silent in the cold. CHECK AGAIN: nothing was judged (a loose wire, no answer) — reconnect and click again.

Accept a box with its own probe (rare)

Only when the box ships with this probe attached. The probe's own calibration is written and the probe must have cleared every bench station.