FloraPulse

Two jobs for the SDI-12 uploader

For researchers. 3 min.

Chapters

  1. 0:00 Two jobs for the SDI-12 uploader
  2. 0:13 What you need
  3. 0:29 Color to color
  4. 0:42 New firmware for a box
  5. 0:54 It runs by itself
  6. 1:13 CONFIRMED means done
  7. 1:26 Any other ending
  8. 1:43 Pair a box with another probe
  9. 1:56 Look up the new probe
  10. 2:13 Upload, then Verify
  11. 2:26 If Verify says anything else
  12. 2:40 A box travels with its probe

What the film says

Two jobs for the SDI-12 uploader

A FloraPulse SDI-12 box holds the calibration of one probe, the one named on its sticker. Two jobs keep a box right, and both run from a page in your browser.

What you need

Both need the small USB uploader, and Chrome or Edge on a computer. One job puts the current firmware on a box. The other loads a probe's calibration, when a box is paired with another probe.

Color to color

Clip the uploader to the box color to color, red to red, white to white, black to black. For a firmware update, take the box off its logger first.

New firmware for a box

For firmware, open the SDI-12 updater page. The first time, click Update box and pick the uploader's port. After that, the page connects by itself.

It runs by itself

Clip a box on. The page checks it and says how long the update will take. That is a minute or two on a recent box, and up to a quarter of an hour on an older one. Then it sends the firmware and confirms it. Keep the tab in front, and everything plugged in.

CONFIRMED means done

CONFIRMED means done. The box keeps its calibration, and it can go back on its logger. Unclip it, and the next box you clip on starts by itself.

Any other ending

If it ends on anything else, leave everything connected and run it again. An update that is never confirmed is rolled back by the box itself. A box shipped before October 2026 cannot be updated this way, and nothing on it is changed.

Pair a box with another probe

To pair a box with another probe, open the calibration upload page and click Connect Uploader. The page reads the box, and a tag shows which probe it holds.

Look up the new probe

Type the new probe's name from its white label, and click Look Up. The calibration comes from FloraPulse, so the computer needs the internet. A red dot beside two names means the box was paired with another probe.

Upload, then Verify

Upload Calibrations writes it into the box. Then Verify reads it back, and takes a reading. With the probe still in its capsule of water, it reads close to zero.

If Verify says anything else

If Verify says anything else, check the clips and that the probe is still in its capsule, then upload again. When it reads good, write the new probe's name on the box.

A box travels with its probe

A box always travels with the probe whose calibration it holds.