What the film says
Testing a sensor
A sensor that has come out of a tree can look dead and be healthy. So before you replace it, test it on the bench. The sensor reuse manual has two checks.
The sensing face
The front of the microchip is the sensing face. Keep it wet, in deionized water. Tap water can foul its pores. Wear gloves, and keep grease off the face.
Clear the face first
If the sensor came out of a tree, clear the old mating compound off the face first. A few minutes in deionized water softens it. Then rinse it gently with a stream of deionized water. Never scrub or wipe the face.
Zero in water
Check one. Stand the sensor in deionized water, connected to its logger, for fifteen minutes to an hour. It should read close to zero, within about one bar, and hold steady.
If the reading is still creeping toward zero, the sensor is not fully wet yet. Wait until the line is flat. Big jumps or noise mean the electronics are likely broken.
The dry-out test
Check two is the dry-out test. First soak the sensor in deionized water for at least an hour. You also need a reading that refreshes at least every five seconds.
The Response Test page
For an SDI-12 sensor, the Response Test page does this for you. Connect the sensor's box to the USB uploader, and open the page in Chrome or Edge. Click Connect Uploader, then Start Response Test. The page asks which sensor it is, so the run is kept under its name.
Settle in water
The page first waits for the reading to settle in the water. That takes about half a minute.
Dry out
Then it asks you to dry the chip. Lift the sensor out, chip facing up, and gently dab only the top of the chip with a clean tissue. Then hands off, and watch the line.
Back in water
As the chip dries, the line climbs. When it is two millivolts above where it started, the page beeps. Put the sensor straight back in the water. Left out, it can cavitate within a minute or two.
Result
A healthy sensor reaches that line within about three minutes. Back in the water, its reading falls by one millivolt within thirty minutes. The page then says sensor passed, and gives you a reference for the run.
A flat line
Three other results can appear. The first is a flat line. After two minutes with no change, the page asks you to check that the chip is out of the water and dry. After five minutes, it reports no response. If you had only just dried the chip, run the test again. If the line stays flat, the sensor is likely cavitated, or its pores are broken.
A slow climb
The second is a slow climb. This sensor took more than four minutes from the start of its climb to reach the line, so its pores are likely partly blocked. It still goes back in the water at the beep, and the page then reports a slow response.
Sand it lightly
Light sanding may clear them. Keep the chip wet, and sand its face with a clean sanding stick, lightly, for sixty seconds. Rinse it, rest it at least thirty minutes in deionized water, and test again. Sand once more if it is still slow, and no more than that.
Up, then straight down
The third climbs, then drops straight back down. That sensor has cavitated, and sanding cannot fix it. Refilling can bring a sensor back, but it takes a pressure station and does not always work. Before you replace it, write to us and quote the reference from the test.
Every step, in the sensor re-use manual
A sensor that passes both checks can go back in a tree. Keep its chip wet until then. Every step is in the sensor reuse manual. The address is on screen.