For advanced users — removing, cleaning, storing, refilling, testing and re-installing FloraPulse probes
This guide provides written instructions for removing and re-using the FloraPulse probes. We recommend watching the videos below for a complete overview of the process.
Links to installation videos

The microchip front is the sensing face. The metal tail is the body you may grip with pliers. The microchip front must be kept wet and away from contaminants throughout removal, storage and re-install.
Store sensors in DI (deionized) water — never tap water. Tap water carries dissolved minerals, chlorine and microbes that can foul the nanopores in the sensing face. Wherever this manual says to keep, soak or store the sensor in water, use DI water. Tap water is fine for cleaning the install site and loosening the cap threads, where it does not reach the sensor face.
FloraPulse sensor installations are generally permanent, but it is possible to remove, store and reinstall sensors. Keep in mind the sensors are fragile and can break during removal. It is best to only move sensors when necessary, and to do so very carefully.
There are 3 main reasons for removing a sensor — here is what we recommend for each.

Sensor removal tools

Extra sleeves, capsules and parafilm come in your install kit. After install, reseal the mating compound to keep it from drying out. These supplies are used again when you reinstall, so keep track of them.
![]() | 1. Carefully cut down the surrounding insulation/wiring. |
![]() | 2. Using a napkin, clean off the excess grease and wounding gels from the install site. A clean area will help prevent contamination of the probe upon removal. You may apply water to help clean the area. |
![]() | 3. Using your hands, twist off the cap. Squirting water on the threads can help loosen the cap. If necessary, use a vise grip to remove the cap. Keep the cap for the sensor re-installation. Fill an extra capsule with DI water and set aside. The capsule should be filled to below halfway, leaving space for the sensor and an air gap. |
![]() | 4. Once the cap is off, rinse the sleeve interior with water to loosen the dried mating compound. |
![]() | 5. To remove the probe, try pulling on the wires by hand (gently, or the wires will break) to see if the sensor is loose and will come out easily. If the sensor is stuck, use needle-nose pliers to gently pull on the metal 'tail' base of the probe and remove the sensor. Do not pull on the wires with pliers. |
![]() | 6. Pull the sensor straight out, and quickly place the sensor inside the plastic capsule with DI water. You may want to rinse the sensor with a stream of DI water before capping. Keep the sensor front wet and in DI water as much as possible. |
![]() | 7. Some installs, especially older ones, can get very dry and stuck. If the sensor still won't come out, fill a container (in this case a Ziploc bag) with DI water. Using your vise clamp, grip the sleeve and pull the sensor and sleeve straight out together. Try to avoid moving/wiggling the sleeve, because this places stress on the probe microchip. Place the sleeve and sensor inside the DI-water-filled container and keep it in DI water until the sensor can be separated. Water hydrates the dried mating compound and the sensor will come loose. |
![]() | 8. Leave the sensor in DI water to loosen the excess powder, then gently rinse it using a stream of DI water. |
![]() | 9. After the sensor is removed and cleaned, cap your sensor and put it aside. You may choose to label the sensor with the datalogger #, port # or anything else to help keep track of where the probe should be reconnected if needed. Sensors must be kept in DI water (not tap water) at all times. |
If a probe cavitates (suddenly jumps to ~0 bar and stays there), you can often bring it back by refilling it under pressure. The steps below are simple and focus on avoiding air and keeping everything wet.
These off-the-shelf parts let you build a small, water-only pressure chamber that connects to a hand hydrostatic test pump:
Safety notes (read first)

Before you start
Keep the sensor wet. If a cavitated sensor dries, nanopores foul over time and will not refill properly. Store the probe front in DI water (not tap water) at all times during handling.
1) Flush the system. With the valve open, run DI water through the pump, hose and vessel to push out all air bubbles. Top off the vessel with water.
2) Bag the probes. Place probes in a well-sealed Ziploc® bag filled with DI water. Carefully push out all air so the bag contains only water and the sensor — no bubbles.
3) Load the chamber. Lower the water-filled bag into the vessel already full of water. Overfill slightly so water beads at the rim, then close the top cap and clamp so no air is trapped.
4) Pressurize. With the valve open, slowly pump to ~600 psi (41 bar).
Pressure ceiling. 600 psi is the target and there is no benefit to going higher. Treat 50 bar (725 psi) as a hard ceiling — that is also the maximum rating of the pump specified above. Probes have been taken to 50 bar without damage, so a brief overshoot is not a reason to discard them, but do not work above it.
5) Close the valve and hold. Close the shut-off valve and leave the chamber at pressure for at least 12 hours (overnight is typical). With the valve closed, the chamber holds its own pressure and a bleeding pump check valve cannot drain it.
6) Check that it held, then depressurize. In the morning, open the valve slowly and watch the pump's gauge:
Release the remaining pressure slowly, open the chamber, remove the bag, and take out the probes.
7) Keep the probes wet. Immediately cap each probe in a DI-water-filled capsule to keep it hydrated. Fill the capsule ~25% with DI water (leaving room for the probe and a small air pocket), insert the probe gently, then blow the water off the connector so the contacts are dry.
Before reinstalling, you can do a short bench check:
These checks mirror the lab tests below, so you can verify function before going back to the field.
If desired, you may perform two tests for each sensor to verify proper functioning.
Clear the sensing face first. Only run these tests after the mating compound has been cleared from the microchip front. Compound left on the sensing face fouls the nanopores and makes the result meaningless — a stalled dry-out trace then cannot tell you whether the pores are partially blocked or the sensor has cavitated.
This takes a few minutes, not an overnight wait: leave the probe in DI water for a few minutes to soften the dried powder, then rinse gently with a stream of DI water. Never scrub or wipe the face. Residue on the metal body, the tail or the wires does not matter and never needs cleaning.
![]() | 1. Place the sensor in a container of DI water for at least an hour to let it relax (fully hydrate) and measure zero pressure. This is the offset voltage (voltage in water). |
![]() | 2. Measure the sensor's drying-out response to confirm the sensor is full of water (not cavitated). For this test you must have a real-time readout of the sensor that refreshes at least every 5 seconds. The readout can be done by connecting the probe to a scientific logger (e.g. Campbell CR6, CR1000). Easier for SDI-12 sensors: no scientific logger is needed — connect the sensor to the FloraPulse SDI-12 interface box and USB uploader, open app.florapulse.com/calibration-upload in Chrome or Edge, and click Start Response Test. The page reads the sensor once per second, plots it live, guides you through this exact test, and gives a verdict automatically. Starting voltage (probe in water) is approx. -1.3 mV. |
![]() | 3. Pull the sensor out of water and hold it with the chip facing up. Gently use a clean napkin to dry off the top of the chip, being careful not to contaminate it, then watch the sensor output. Wear protective gloves to avoid contaminating the sensor. The sensor will not work if grease or contaminants get on the pores (top of the chip). |
![]() | 4. The sensor output voltage will soon shoot up and start reading increasing tension. Wait until the sensor reads -10 bar (about offset + 2 mV) and then immediately place it back in water. This process should only take a few minutes. It is critical to watch the sensor output carefully, because once the sensor starts drying out it will cavitate within a minute or two if not quickly placed back in water. Ending voltage returned to -1.3 mV after being returned to water. |
The table below shows the 3 stages of the response test (measure offset, dry sensor, re-hydrate sensor), the expected 'good' behavior for each, and potential problems and solutions.
| Stage | Good | Problems | Solutions |
|---|---|---|---|
| Measure offset | Sensor offset should be flat and stable. | Large noise or random jumps. Measured offset slowly decreasing. | The sensor electronics are likely broken. Use a new sensor. The sensor is still not fully re-hydrated. Wait until the line flattens out. |
| Dry sensor to offset+2 mV (-10 bar) | Sensor output shoots up to offset+2 mV in a straight line within ~3 minutes. | Sensor output does not change during dry-out (stays flat). Sensor output increases very slowly and takes 3+ minutes to reach +2 mV. Sensor output increases, then drops sharply back to offset. | Sensor is cavitated or the nanopores are broken. Use a new sensor. Sensor is partially blocked. Re-hydration should also be slow. Try sanding the pores to increase response speed (see below). Sensor cavitated because it dried too much (too much tension), or the pores broke. Use a new sensor. |
| Re-hydrate sensor | Sensor output decreases back towards offset. Voltage goes down by 1 mV within 30 minutes. | Sensor output decreases very slowly, takes 30+ minutes to decrease by 1 mV. Sensor output drops sharply back to offset. | Sensor is partially blocked. Try sanding the pores to increase response speed (see below). Sensor cavitated. Use a new sensor. |
Sometimes sensors are slow due to contamination, and it may be possible to remove this contamination by lightly sanding the pore surface. This process should only be done carefully and only when necessary, because excessive sanding will eventually break the sensor.
![]() | Sand the chip surface. Pull the sensor out of water and leave the pore surface wet to prevent dry-out and cavitation. Using a clean sanding stick, apply light pressure and abrade the chip top surface back and forth for 60 seconds, at a rate of 2 strokes per second. You should hear a scratching noise as the sanding stick abrades the nanopores to open them up. It may be necessary to rotate the sanding stick as the abrasive wears out, to use fresh abrasive. |
![]() | Rinse the sanding debris. Place the sensor back in DI water and swish it around to rinse off sanding debris. |
Let the sensor rest for 30+ minutes in DI water to fully re-hydrate, then re-test for response speed as explained above (Test #2).
You may repeat sanding one more time if response speed is still slow. Further sanding will not help and will eventually destroy the sensor.
Slow sensors will still measure water potential, but their response will lag the tree water potential, so accuracy will be impacted.
To reuse the sensor, you will need to gather the following materials:
See the installation manual for the full list of install tools and installation instructions.